Specifications for the TA044 differential probe:
- Bandwidth: DC to 70 MHz (-3 dB)
- Attenuator: 100:1, switchable to 1000:1
- Input impedance: 10 MΩ / 10 pF from each input to ground
- Input voltage:- Max. differential voltage at 1/100 attenuation: ±700 V (DC + AC peak) or 700 V rms- Max. differential voltage at 1/1000 attenuation: ±7000 V (DC + AC peak) or 5000 V rms—Common-mode input voltage: ±7000 V (DC + AC peak) or 2500 V rms (at 1/100 and 1/1000 attenuation)
- Maximum absolute input voltage: - Differential or common-mode: ±7000 V (DC + AC peak) or 2500 V rms CAT III (at 1/10 and 1/100 attenuation)
- Accuracy: ± 2%
- Output voltage:- Signal (into a 50 kΩ load): ±7 V- Offset (typical): < ± 5 mV- Noise (typical): 0.9 mV rms
- Source impedance (typical): 50 Ω
- Supply voltage: 4 x AA batteries or 6 V / 200 mA DC via 9 V DC / 120 mA DC adapter
- Length of connection cables: 60 cm
- Length of shielded BNC cable: 90 cm
- Weight: 500 g
- Dimensions: W 83 x L 202 x H 38 mm
Please note: Special-stock items may show signs of use. Subject to prior sale. The configuration of a special-stock item cannot be changed. The calibration of a special-stock item may no longer be up to date. If you would like a current calibration, please feel free to contact our experts.
Give us a call— we’d be happy to advise you personally:
+49 (0)7121 / 51 50 50